F. Basan et al.
At the pontoon site in the Stockholm archipelago, projector and recorder were
deployed at 5 m depth with a horizontal separation of 5.6—-8.3 m. Here, end-wall
reflections are irrelevant, but the surface reflection curve becomes the dominant con-
straint. With the larger depth and separation, usable frequencies extended down to
800 Hz (Fig. 1).
Signal Design and Analysi:
Projector signals were emitted as short narrowband pulses and repeated three times
per frequency step. Sensitivities were obtained by matched-filter detection of each
pulse and comparison of the resulting amplitudes with the absolute pressure derived
from the projector’s TVR and drive voltage. To reduce variance, the three repeats
were averaged per frequency.
Results were averaged to decidecade bands based on definitions in ISO
18405 (2017).
The calibration exercises covered representative autonomous recorder systems used
in Baltic Sea monitoring programs. Table 1 lists the devices tested in the FOI indoor
vasın (2-10 kHz), including their input voltage ranges. Table 2 provides the
corresponding overview for the open-sea pontoon calibration (0.8-20 kHz). The spec-
ification of voltage ranges is particularly relevant, as it determines the scaling of the
digital counts to voltage and thus allows calibration results to be expressed relative to
full scale as clip level, 1e., the maximum level the system can record before the audio
files clip.
During the second day at the pontoon, several units did not record as intended (e.g.,
logging/power supply issues), so fewer systems were successfully calibrated. As a
result, one-to-one comparisons with the tank results are only possible for the overlapping
devices.
Table 1 Overview of recorders that were calibrated in the FOI tank during the first day of the
HELCOM calibration workshop
Country/institution
Sweden/FOI
Sweden/FOL
Sweden/FOL
Germany/BSH
Germany/BSH
Denmark/AU
Estonia/TUT
Recorder
Soundtrap ST640
Soundtrap ST300HF
Sylence LP.
Sylence LP.
LIE.
Soundtrap ST600HF
Sylence LP
Hydrophone
HTI-96-UHF
Insitu
HTL92-WB
HTI-96-min
In situ
In situ
HTI-92-WB
; Voltage range
+1 V
+1 V
+2.5 V
+2.5 V
+3 V
+1 V
"45V