Cruise Report, RV ATAIR, Cruise 067I033-1, Bremerhaven – Thybor?n – Aberdeen, 23 Jul – 15 Aug 2024 17
In summary, any post-cruise correction of the data was not applied, and the temperature, con-
ductivity and salinity data were made available with flag 1 (good data). The usage of the data
from the secondary sensor pair (TC2) is recommended.
Sampling and titration of oxygen samples was not performed during cruise 067I033-1 to the
North Sea, but was carried out on the subsequent cruise 067I033-2 to the German Bight. The
same oxygen sensor was used on both cruises. The comparison between CTD oxygen and
oxygen obtained from titrated water samples of cruise 067I033-2 unfortunately revealed a mal-
function of the CTD oxygen sensor. Consequently, any correction of the sensor-derived oxy-
gen data was not possible, and sensor-derived oxygen data from both cruises were made
available with flag 4 (bad data).
Any quality control of fluorescence and turbidity data was not performed. Thus, the respective
data were made available with flag 0 (no quality control applied).
Salinity Difference
AUTOSAL-CTD
All samples
All samples with
differences < +/- 1
standard deviation
All samples with
differences < +/- 1
standard deviation
for the 2nd mean
AUTOSAL - TSC1
Number of data pairs 183 175 150
percentage 96% 82%
Mean difference -0.0074 -0.0008 -0.0007
Median difference -0.0012 -0.0011 -0.0010
Standard deviation 0.0466 0.0059 0.0021
AUTOSAL – TSC2
Number of data pairs 183 175 111
percentage 96% 83%
Mean difference -0.0064 -0.0007 -0.0005
Median difference -0.00008 -0.00004 -0.00008
Standard deviation 0.0460 0.0065 0.0020
Table 5.3. Results of the comparison between bottle salinity samples obtained from salinometer meas-
urements (AUTOSAL) and conductivity/salinity recordings at the time of closing the Niskin bottles, the
latter obtained from the two CTD sensor packs (TSC1 and TSC2).