20 | Cruise Report, RV ATAIR, Cruise 0671021-1, Bremerhaven - Thyboron —- Aberdeen, 18 Jul - 10 Aug 2023
values of 5-6 ml/l, this threshold corresponds to 0.055 mI/l. Without correction, the average
difference of the oxygen data is 0.0967 ml, Le. greater than 1%; with correction, it is well
below 1%. CTD oxygen was corrected according to:
Deor (MI/I) = -0.888808751227 + 0.003882681122 * time + 0.021469570016 * oxygen (ml/l) +
-0.000104482816 * pressure + 0.002460619138 * temperature
Oxygen saturation was recalculated afterwards. The corrected CTD-derived oxygen data was
given the quality flag 1 (good data). Any quality control of fluorescence and turbidity data was
not performed. Thus, the respective data were made available with flag 0 (no quality control
applied).
STABILE:
AT
“x U 911
AyQll
17
Se EC EEG
ES A
CE
En deviatio
A mean
AUTOSAL - TSC1
Number of data pairs 1
percentage
Mean difference
Median difference
Standard deviation
JE
.0.004*
-0.0015
0.0358
"5
rl
99 5%
0.0016
-0.0014
0.0079
1C5
824%
-0.0006
-0.0009
0.0036
AUTOSAL —- TSC2
Number of data pairs
percentage
Mean difference
K
3
1
-0.0030
——
Median difference -0.0010
Standard deviation
0 02361
123
98%
-0.0007
-0 0010
00072
161
82%
-0.0005
-0.0010
0 00392
Table 5.3. Results of the comparison between bottle salinity samples obtained from salinometer meas-
urements (AUTOSAL) and conductivity/salinity recordings at the time of closing the Niskin bottles, the
latter obtained from the two CTD sensor packs (TSC1 and TSC2).