2 |
Cruise Report, RV ATAIR, Cruise 0671021-1, Bremerhaven - Thyboron —- Aberdeen, 18 Jul - 10 Aug 2023
PSA-916. Inlets of the two T/C sensor pairs were soaked in distilled water before the start and
after the end of each CTD cast. A supplementary temperature sensor was installed at the
height of Niskin bottle about 1 m above the CTD unit in order to record water temperature at
the time of the bottle closing. All temperature and conductivity sensors were successfully re-
calibrated in the BSH’s calibration laboratory prior to the cruise. All devices and sensors that
were in use are listed in Table 5.1, the Niskin bottle set-up is listed in Table 5.2. Temperatures
are reported on the /TS-90 scale. Derived salinities refer to the Practical Salinity Scale of 1978
(PSS-78). All respective densities have been inferred from the EOS-80 equation.
DE
CTD main housing with pressure sensor,
SBE9plus
deckunit, SBE11plus
temperature sensor 1, SBE3T, part of TC1
temperature sensor 2, SBE3T, part of TC2
conductivity sensor 1, SBE4C, part of TC1
conductivity sensor 2, SBE4C, part of TC2
oxygen sensor, SBE43
pump 1, SBE5
pump 2, SBE5
dual-wavelength sensor, WET Labs ECO-AFL/FL,
fluorescence
dual-wavelength sensor, WET Labs ECO, turbidity
supplementary termperature sensor installed at
the height of the Niskin bottles, SBE35
altimeter, Teledyne Benthos. PSA-916
carousel water sampler with 12 positions for Niskin
bottles SBE23?2
EINE
09P21787-0577, S1
11P29178-0620. N1
5290
478
37792
2752
0171
5339
5856
4 - 4
Ar
ACC»
O3
51678
3232948-0466. K1
Calıbration Date
06 Feb 2019
29 Jun 2023
29 Jun 2023
29 Jun 2023
29 Jun 2023
16 Apr 2022
15 Jan 2018
15 Jan 2018
04 Jul 2016
11 Dec 2013
Table 5.1. Overview on CTD-related sensors or water sampler devices used during cruise 0671021-1.