38
BSH Report No. 56
A look at the hardware
GC-FID-
BSH
IAEA
NFI
RWS
Conditions
Hardware and Operation
Instrument
Agilent 7890A
Agilent 7890A
Agilent 6890N
Agilent 7890A
Detector
FID
FID
FID
FID
Carrier gas
He, H2, other
H2
He
N2
He
Injection
Volume [pL]
1
1
1
1
Injection temp. [°C]
320
300
300
325
Pre-lnject. delay [sec]
1
0
0
5
Injector type
solvent vent*
spiltless
split, 20:1
splitless
Column
Name
J&W DB-5
HP-5
HP Ultra-1
J&W DB-1
Length [m]
10
30
25
15
Diameter [mm]
0.1
0.25
0.2
0.25
Film [pm]
0.1
0.25
0.33
0.25
Const, flow [mL/min]
0.92
1.2
0.4
1.1
Oven program
Start [°C]
40
60
60
35
Hold_start [min]
0.71
1
1
1.5
Rate_l [°C/min]
30.2
6
30
10
Rate_l_end [°C]
300
310
150
325
Hold_l [min]
2.5
-
0
-
Rate_2 [°C/min]
15
-
5
-
End [°C]
325
310
320
325
Hold_end [min]
3.6
15
12
10
Run time [min]
17.0
57.7
16.0
40.5
FID-Conditions
Heater [°C]
340
300
250
375
H2 [mL/min]
30
35
30
40
Air [mL/min]
320
350
320
300
Makeup [mL/min]
30
28.09
30
25
*= The "Solvent Vent Mode" is a setting unique to PTV-injectors. It is a special type of splittless injection but allows the handling of large
volumes when needed.
As can be seen from the table above, GC-FID conditions vary a lot between
the participants. Carrier gas, column type and separation conditions can there
for contribute to the systematic differences observed.