- 8 -
7; Hardware components of SEAS-II unit and near real-time
data flow
As XBT probes Sippican "Deep Blue" only were launched. The
specification given by the manufacturer is as follows:
Temperature:
Depth:
Range
Accuracy
Resolution
-2 °C to 35.5 °C
± 0.2 °C
to 0.01 °C
Range
Accuracy
Resolution
to 760 m
± 2 % of depth
65 cm
Owing to the fact that the probes were deployed from the wing
of the navigation bridge (height: 30 m) a depth accuracy in
fluencing effect should be expected (Krause, 1986), however,
its quality cannot be given.
To date, no sea surface temperatures are measured indepen
dently aboard CMS "Koln Atlantic" as a reference. Thus, for
the WOCE field phase it is hoped to install a thermometer
within the ship at the ship's side, equipped with a display
on the navigation bridge. Furthermore, a vertical extension
of measurements down to 1500 m is planned.