Volltext : Oil Spill Identification - Round Robin 2005

18

Oil  Spill  Identification  -  Round  Robin  20045

peaks.  This  might  be  concentration  related:  in  combination  with  a  small
(ID  0.18  mm)  column  overloading  might  be  a  reason  for  this  behavior.
The  MS  chromatogram  of  Source  II  is  also  strange.  Resolution  is  far  less
compared  to  the  other  chromatograms.  As  a  result  Source  II  has  been
excluded.  Comparing  Fig  3  and  Fig  4  of  the  BMM  report  indicates  that
the  injection  concentration  between  source  and  spill  samples  is  a  factor
of  10  or  more.  It  is  advised  to  optimize  the  injection  concentration.
The  evaluation  of  the  chromatograms  is  laborious  and  susceptible  to
failures.  Patrick  has  indicated  that  he  intends  to  improve  data  handling
Differences  between  the  ratios  of  small  peaks  can  be  larger  than  normal
because  of  the  difference  in  the  injection  concentration.  This  is  e.g.
visible  in  the  m/z  191  and  m/z  216  chromatograms.
GC-screening  may  be  used  to  adjust  the  concentration  of  samples,  as  it
is  done,  for  example,  by  SKL:
After  a  first  preliminary  GC-FID  run  we  eventually  dilute  or  concentrate  the  extracts  so
they  all  will  have  approximately  the  same  concentration.  Most  of  our  samples  that  are
collected  from  the  water  are  taken  with  a  teflon  cloth,  so  the  concentration  can  vary  a
lot.

2.2  Centre  de  Documentation  de  Recherche  et
d'Experimentation  sur  les  pollutions  accidentelles  des
eaux  (Cedre)
Contact:  Julien  Guyomarch
Cedre  (http://www.le-cedre.fr/1  was  created  in  1978  within  measures
taken  after  the  wreckage  of  the  "Amoco  Cadiz",  to  improve
preparedness  against  accidental  water  pollution  and  strengthen  the
national  response  organization.  Its  expertise  encompasses  both  marine
and  inland  waters.
The  RR-samples  have  been  analyzed  with  GC-MS  in  SIM  mode  on
alkanes,  PAH's  and  biomarkers.
=>  The  comparison  of  the  samples  Source  I  and  II,  and  Extract  I  and  II  led  to
the  following  conclusions:
-  n-alkanes  show  differences  in  the  region  n-Cl  1  to  n-C19,  but  these
variations  are  not  significant  due  to  possible  evaporation  processes  affecting
these  distributions.  On  the  other  hand,  the  relative  abundances  of  compounds
ranking  from  n-C20  to  n-C30  did  not  show  differences.
-  The  biomarkers  (m/z=191,  217,  218,  231)  did  not  allow  to  differentiate  the  4
samples  due  to  high  variability.
-  Finally,  the  diagnostic  ratios  calculated  from  PAHs  analyses  present
variations  that  could  explain  differences  of  origins  for  Source  I  and  II.  These
significant  differences  showed  for  fragment  216  and  234.
=>  The  conclusion  of  this  oil  spill  identification  is  that  the  samples  collected
at  the  water  surface  do  not  seem  to  come  from  the  two  boats.  Moreover,  the
two  bilge  samples  and  the  two  surface  samples  have  different  origins.
Notes/remarks:
            